Fix Sirona SIDEXIS XG 'Data could not be stored' error during 2D panoramic exposure

draft · imaging sironasidexis

Generated by docuprocessor (prompt article/v3) · 2026-04-22 15:22

Source thread

thread_id: spaces/AAAA05BdS6s/threads/V2YtyHbutAQ · 14 msgs · first 2025-01-28 · participants: Tech ATech B

Tech A 2025-01-28 17:08
for Blue Wave Orthodontics - Rye, when trying to take a 2d Panoramic, Sirona is waiting for Exposure, and they are taking the xrays, but instead we got an error message:
image/png image.png
**Screen:** Sirona dental imaging software dialog titled *"2D exposure: 2, DemoX-ray \*03/19/01"* with a **Preview** pane (blank/no image loaded). **Status message:** *"Device is ready for exposure. Radiation can be triggered."* **Indicator:** A solid **bright green square** is visible in the lower-right of the status bar, consistent with a ready/armed state indicator; a **2D tooth icon** is displayed on the left of the status row. A **Quit** button is present; no error codes or fault indicators are displayed.
Tech A 2025-01-28 17:08
(no text)
image/png Screenshot 2025-01-28 113818.png
**Software:** SIDEXIS XG – 2D Exposure session in progress for "DemoX-ray \*03/19/01" **Error Dialog (XGData):** Message reads: *"Data could not be stored."* Detail text: *"[CXGSession::SmfWfimageblockImageend] Image could not be saved"* **UI State:** A 2D exposure preview thumbnail is visible in the lower-left of the exposure window; the session toolbar shows a red recording indicator ("Session in progress").
image/png Screenshot 2025-01-28 113753.png
**SIDEXIS XG** software (version indicated as "2, DemoX-ray \*03/19/01") is displaying a modal error dialog from the **XGData** module with the message: **"Data could not be stored."** The dialog contains a red X error icon, an **OK** button, and a **Details** button. A secondary window labeled **"2D exposure: 2, DemoX-ray \*03/19/01"** is open in the background showing **"Exposure in progress"** status with a **Quit** button visible; the software is identified as a **Sirona** product.
Tech A 2025-01-28 17:09
They had restarted the machine, and also the PC, user profile has access to PDATA folder.
Tech B 2025-01-28 17:13
is there enough space on the aqc PC?
Tech A 2025-01-28 17:14
110 GB Free
Tech B 2025-01-28 17:18
so, the xray can be taken but not saving, right?
Tech B 2025-01-28 17:19
i've never seen this error before. can u try to import a random image into Sidexis and see if this works? If it doesn't work, then it might be something with the temp folder or pdata settings if it works, then might be the machine itself or drivers
Tech A 2025-01-28 17:19
i would say not even taken, as soon as they push the radiation, the error on Sidexis comes up, not showing any Xray taken just the errors
Tech B 2025-01-28 17:20
then its the machine itself but try to import the image to be100% sure
Tech A 2025-01-28 17:21
it let me import an image
Tech A 2025-01-28 17:21
is a random one i downloaded
image/png image.png
**Software:** SIDEXIS XG – 2, remote session via LogMeIn. **File/Session Info:** DemoX-ray \*03/19/01, image file labeled "Import.test.jpg, Original," timestamp shown as 01/28/25. **Display:** Panoramic dental radiograph (BWO/PAN) is open in the SIDEXIS XG viewer window; image appears fully rendered with no visible error dialogs or warning overlays. **Window Title:** "MARO RYE – PAN (BWO-XRAY-0922) – Remote Control," suggesting patient/unit identifier BWO-XRAY-0922.
Tech B 2025-01-28 17:49
then, its the machine
Tech A 2025-01-28 18:05
Vendor Call?
Tech B 2025-01-28 18:06
either call the vendor or the recycle company 🙂